Submit date: (b)(6) 2016.A device history record review could not be performed because the lot number provided is invalid.There were no samples received for evaluation.Because a sample was not available for evaluation, a root cause analysis could not be conducted to determine the root cause of this reported issue.If samples are received at a later date, this complaint will be re-opened and the investigation continued.A corrective action is not required at this time.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, in process visual inspection and visual acceptance sampling) are in place to prevent non-conforming product from leaving the manufacturing operations.As per procedure, manufacturing performs 100% visual inspection during production, which would identify nicks or cuts in the catheter assembly.This complaint will be used for tracking and trending purposes.
|