MAUDE Adverse Event Report: BIOMET 3I CERTAIN® GOLD-TITE® HEXED SCREW; ABUTMENT SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I CERTAIN® GOLD-TITE® HEXED SCREW; ABUTMENT SCREW Back to Search Results
Catalog Number IUNIHG
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/10/2015
Event Type  malfunction  
Event Description
The dentist reported the gold hex screw broke off at site # 14, in the implant.
 
Manufacturer Narrative
Upon visual inspection, the screw is fractured at approximately the first thread.There is foreign material in the hex of the screw, so the condition of the hex cannot be determined.The device history record for the screw could not be reviewed as no lot number was provided.A definitive root cause has not been determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN® GOLD-TITE® HEXED SCREW
Type of Device
ABUTMENT SCREW
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
dania perez
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key5062166
MDR Text Key25420387
Report Number0001038806-2015-00793
Device Sequence Number1
Product Code NHA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK072642
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Report Date 08/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberIUNIHG
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/09/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age44 YR
-
-