MAUDE Adverse Event Report: INTEGRA YORK, PA INC. RHOTON-TYPE HOOK 71/2 90D SSHP; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA YORK, PA INC. RHOTON-TYPE HOOK 71/2 90D SSHP; N/A Back to Search Results
Catalog Number 277268
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/27/2015
Event Type  malfunction  
Manufacturer Narrative
To date the device involved in the reported incident has not been received for evaluation.An investigation has been initiated based on the reported information.
 
Event Description
Customer initially reports surgeon was performing an anterior cervical fusion.He noticed that the instrument (a rhoton dissector) was missing a part of its tip.X-ray taken at end of procedure was negative for presence of foreign body in the incision site.No harm to patient.
 
Manufacturer Narrative
(b)(4).Integra investigation completed.Manufacture date unknown.Method: failure analysis, device history evaluation results: failure analysis - there is one rhoton-type hook returned in used condition, showing minimal wear.Inspected the hook and the tip is broken off, this is a very delicate instrument and could have happened while in use or even dropped.This complaint is confirmed/root cause undetermined.Device history evaluation - nonconforming product report / nonconforming material report history: there is no applicable nonconforming product report / nonconforming material report history.Variance authorization / deviation history: none.Engineering change order/manufacturing change order history: there is no applicable engineering change order/manufacturing change order history.Corrective action preventive action history: none.Health hazard evaluation history: none.Conclusion: the root cause has not been identified as a workmanship or material deficiency.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RHOTON-TYPE HOOK 71/2 90D SSHP
Type of Device
N/A
Manufacturer (Section D)
INTEGRA YORK, PA INC.
589 davies drive
589 davies drive
york PA 17402
Manufacturer (Section G)
INTEGRA YORK, PA INC.
589 davies drive
york PA 17402
Manufacturer Contact
sandra lee
311 enterprise drive
plainsboro, NJ 08536
6099362393
MDR Report Key5062384
MDR Text Key25229505
Report Number2523190-2015-00075
Device Sequence Number1
Product Code GZX
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Health Professional
Report Date 08/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number277268
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/25/2015
Initial Date FDA Received09/08/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age46 YR
-
-