Model Number SC-1110-02
Device Problem Telemetry Discrepancy (1629)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/03/2015
Event Type  malfunction  
Event Description
A report was received that the patient's remote control would not connect to the patient's ipg.It was also reported that the patient had a nondevice related surgery wherein electrocautery might have been used.The patient was referred to a surgeon for a battery replacement.Monopolar electrocautery is a known source of high voltage transient signal and current company labeling warns against the use of monopolar electrocautery (physician's implant manual 9055940-001).
 
Manufacturer Narrative
Additional information was received that there will be no further course of action at this time.The device was not returned to bsn.It is indicated that the device will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed.A review of the device history records will be conducted.If there is any further relevant information from that review, a supplemental med watch will be filed.
 
Event Description
A report was received that the patient's remote control would not connect to the patient's ipg.It was also reported that the patient had a nondevice related surgery wherein electrocautery might have been used.The patient was referred to a surgeon for a battery replacement.Monopolar electrocautery is a known source of high voltage transient signal and current company labeling warns against the use of monopolar electrocautery (physicianimplant manual 9055940-001).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION®
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key5062452
MDR Text Key25216031
Report Number3006630150-2015-02280
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Followup
Report Date 08/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/30/2009
Device Model NumberSC-1110-02
Other Device ID NumberM365SC1110020
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/08/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/26/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/13/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-