1) this is a re-filing of report 3008289288-2014-00004 with a new report number.The report was edited on (b)(6) 2015 to revise the facility registration number from (b)(4).The current facility registration number changed in may 2014 when natus neurology inc.Moved to its current location at (b)(4).Patient information was not provided, patient age and weight are therefore defaulted to "0".Device evaluated by manufacturer: device was not returned to manufacturer for evaluation.Registered internally as a complaint ((b)(4)) for further investigation by manufacturer.The internal investigation could not reproduce the problem condition reported nor did the complainant respond to further information requests.Subsequently, no other follow up actions were taken.
|
While performing an eeg exam using a nicolet cortical stimulator, the healthcare professional using the stimulator noted that the equipment reported an output stimulation level of "0" when stimulation was applied even though the equipment was set to deliver stimulation.A small level of stimulation was also noted as an eeg artifact although the level appeared low when compared to other equipment later used (once noticed, the healthcare professional switched to a different stimulation method).
|