A user facility reported during initiation of treatment, blood was seen leaking from underneath the arterial port.The nurse suspected a loose connection may have caused the leak.Blood was not returned to patient and the estimated blood loss was 200 cc's.There were no adverse effects to the patient.The blood flow rate was documented at 375 ml/min.The blood lines were replaced with a new set, and patient was able to successfully complete their treatment.
|
Patient's weight was entered incorrectly.(b)(6).The device was not returned to the manufacturer for physical evaluation and field service was not requested, therefore, the failure mode cannot be confirmed.The facility biomedical technician confirmed he could not duplicate the issue.However, an investigation of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, the batch record review confirmed the labeling, material, and process controls were within specification.
|