Method: visual inspection; functional inspection; device history review; complaint history review; risk assessment; results: the stryker rep confirmed that some holes were prepared with the drill guide and some weren't, which is the probable cause of the spring bar deformation.The surgical technique recommends the use of the drill guides as it provides the correct angle of trajectory so as not to damage the plate.Conclusion: the root cause is deviation from the recommended technique outlined in the surgical technique.
|