• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-FRANCE AVIATOR ASSY ONE LEVEL PLATE SIZE 12; SPINAL INTERVERTEBRAL BODY FIXATION ORTHOSIS.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-FRANCE AVIATOR ASSY ONE LEVEL PLATE SIZE 12; SPINAL INTERVERTEBRAL BODY FIXATION ORTHOSIS. Back to Search Results
Catalog Number 48811112
Device Problems Material Separation (1562); Material Deformation (2976); Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/23/2015
Event Type  malfunction  
Event Description
It was reported that; spring bar lifted on a one level aviator plate.
 
Manufacturer Narrative
Method: visual inspection; functional inspection; device history review; complaint history review; risk assessment; results: the stryker rep confirmed that some holes were prepared with the drill guide and some weren't, which is the probable cause of the spring bar deformation.The surgical technique recommends the use of the drill guides as it provides the correct angle of trajectory so as not to damage the plate.Conclusion: the root cause is deviation from the recommended technique outlined in the surgical technique.
 
Event Description
It was reported that; spring bar lifted on a one level aviator plate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVIATOR ASSY ONE LEVEL PLATE SIZE 12
Type of Device
SPINAL INTERVERTEBRAL BODY FIXATION ORTHOSIS.
Manufacturer (Section D)
STRYKER SPINE-FRANCE
zone industrielle de marticot
cestas 33610
FR  33610
Manufacturer (Section G)
STRYKER SPINE-FRANCE
zone industrielle de marticot
cestas 33610
FR   33610
Manufacturer Contact
christa marrow
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key5070470
MDR Text Key26073752
Report Number0009617544-2015-00399
Device Sequence Number1
Product Code KWQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142237
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Followup
Report Date 06/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number48811112
Device Lot Number147460
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/11/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/21/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/26/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-