Model Number 9-ASD-036
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Embolism (1829)
Event Date 08/20/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
This 36 mm amplatzer septal occluder (aso) was implanted; however, after release from the delivery cable, the aso embolized.The aso was removed via a femoral vein dissection using a snare system and a biopsy forceps.
 
Manufacturer Narrative
(b)(4).The results of this investigation confirmed the aso met dimensional specifications when analyzed at sjm.A review of the device history record confirmed the occluder met all visual, dimensional, and functional specifications at the time it was manufactured, prior to shipment.There was no evidence to suggest there was an intrinsic defect in the aso, and the cause for the embolization remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key5081278
MDR Text Key25970895
Report Number2135147-2015-00104
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Followup
Report Date 08/26/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/29/2020
Device Model Number9-ASD-036
Device Catalogue Number9-ASD-036
Device Lot Number5024693
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/09/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/24/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-