The device referenced in this report was returned to olympus for evaluation.The reported phenomenon was confirmed.A physical inspection of the device was performed and it was found that the bending section cover glue was cracked and had been taped.It was also observed that there were excessive frayed wires on the device.Due to the condition of the bending section a leak test could not be performed.This type of damage is most likely due to mishandling or improper storage.The scope was refurbished and returned to the user facility.The instruction manual warns users "the probability of failure of the endoscope and ancillary equipment increases as the number of procedures performed and/or the total operating hours increase.In addition to the inspection before each procedure, the person in charge of medical equipment maintenance in each hospital should inspect the items specified in this manual periodically following regulations, guidelines, etc.Required of you.An endoscope with an observed irregularity should not be used.".
|
Olympus was informed that during a navigational bronchoscopy procedure, the black coating (bending section rubber glue) on the device was noted to be missing.It was reported that no device debris fell inside the patient.The procedure was completed with the same device.There was no patient injury reported.Olympus followed up with the user facility to obtain additional information regarding the reported event and was informed that the scope is manually cleaned; leak tested, and sterilized using a v-pro max on the flexible cycle.This is the facility's only bronchoscope.
|