Catalog Number 10492730
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Code Available (3191)
Event Date 08/15/2015
Event Type  malfunction  
Manufacturer Narrative
Siemens current product service (cps) team investigated the data files and associated lab results and concluded that k+ sensor exhibited acceptable, typical performance.Discrepancy most likely due to pre-analytical factors and differences between ise direct vs.Indirect measurement.
 
Event Description
Customer reported discordant potassium on the analyzer.Customer indicated that unnecessary potassium treatment was given to the patient due to discordant result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key5082638
MDR Text Key26246486
Report Number1217157-2015-00135
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Report Date 09/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Catalogue Number10492730
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/16/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-