The customer reported that they received erroneous results for ten patient samples tested for tina-quant hemoglobin a1c gen.2 (hba1c) on an integra 400 plus analyzer.The customer stated that they had switched to a new cassette of the same reagent lot number when the erroneous results occurred.The customer stated that some of the initial results were accompanied by data flags, but could not provide information on which specific results were accompanied by data flags.All initial results were reported outside of the laboratory where they were questioned by a physician.After the sample results were questioned, the customer ran controls and these controls were out of range.Controls were repeated and these were within range, so then the samples were repeated.The repeat results were believed to be correct.The first sample initially resulted as 9.8 % and the repeat result was 6.2 %.The second sample initially resulted as 9.6 % and the repeat result was 6.5 %.The third sample initially resulted as 9.6 % and the repeat result was 6.2 %.The fourth sample initially resulted as 12.1 % and the repeat result was 8.6 %.The fifth sample initially resulted as 10.5 % and the repeat result was 8.6 %.The sixth sample initially resulted as 9.5 % and the repeat result was 6.0 %.The seventh sample initially resulted as 11.4 % and the repeat result was 7.9 %.The eighth sample initially resulted as 14.5 % and the repeat result was 9.8 %.The ninth sample initially resulted as 10.2 % and the repeat result was 7.2 %.The tenth sample initially resulted as 14.3 % and the repeat result was 9.9 %.The patients were not adversely affected.The integra 400 plus analyzer serial number was 39-9439.The field service representative stated that the customer determined that the reagent cassette was bad.The customer replaced the cassette and calibrated the new cassette.
|