• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 06437281190
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/20/2015
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer received questionable thyroid results for one patient sample from cobas e411 analyzer serial number (b)(4).The sample was submitted for investigation and was tested on a cobas e411 analyzer and a centaur analyzer.The results for thyrotropin (tsh), free thyroxine (ft4), and free triiodothyronine (ft3) were discrepant.Refer to the medwatch with (b)(6) for the other assays.Information concerning if any erroneous result was reported outside the laboratory was requested, but was not provided.No adverse event was reported.A specific root cause could not be identified as insufficient sample material remained for further investigation.From the information provided, a general reagent issue was not likely.When comparing results generated by different types of analyzers, variances can be expected due to the different setups of all assays, the antibodies used and the variances in reference methods.For thyroid parameters, the patient's age, gender, and other characteristics should be considered when analyzing results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5085702
MDR Text Key26115118
Report Number1823260-2015-04189
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Report Date 09/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/30/2016
Device Catalogue Number06437281190
Device Lot Number184927
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Event Location Laboratory
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/17/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age031 YR
-
-