As reported, the patient presented with a very tight venous stenosis.Target treatment site would not respond to a 12mm balloon dilation.A 13mm x 10cm gore viabahn endoprosthesis was positioned at the treatment site.During deployment, the device deployed; however, as the deployment line was removed from the introducer sheath, part of the nitinol wire appeared to be removed as well.Reportedly, the deployment line appeared wrapped around the nitinol wire.The end of the wire was cut releasing the deployment line.A 13mm x 5cm gore viabahn endoprosthesis device was then advanced though the sheath pushing the nitinol wire was pushed back into the sheath.At the end of the procedure, both of the devices appeared patent.The patient was fine post procedure.
|