Model Number 2AF284 |
Device Problem
Insufficient Information (3190)
|
Patient Problems
Low Blood Pressure/ Hypotension (1914); Blood Loss (2597); Pericardial Effusion (3271)
|
Event Date 09/16/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.(b)(4).
|
|
Event Description
|
It was reported that during a cryoablation procedure, while ablating the third vein, the patient's blood pressure dropped to an extremely low status.Ultrasound was checked for pericardial effusion.The ultrasound confirmed that a small pericardial effusion was seen.The patient's abdomen appeared to be filling with fluid/blood.The case was aborted and the patient was taken to the operating room for further examination.
|
|
Manufacturer Narrative
|
Product event summary: the cryo balloon catheter (b)(4) was visually inspected, functionally tested.Bin files were not returned for this event.This is a clinical issue encountered during the procedure.Upon visual inspection of catheter results showed the achieve mapping catheter was stuck on the push button segment of the catheter.Smart chip verification indicated the catheter was used for eight injections.The catheter passed the performance test; electrical integrity and impedance were also within specification.Dissection and pressure testing did not show any leaks or traces of liquid/blood inside the catheter.The catheter passed final inspection.The yield for this lot was within historical yield range for this product.In conclusion, this is a clinical issue encountered during the procedure.The balloon catheter passed the returned product inspection as per specification.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|