MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS OSCILL-SAW-ATTACHM F/532.001 532.010 532; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS OSCILL-SAW-ATTACHM F/532.001 532.010 532; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.021
Device Problems Partial Blockage (1065); Sticking (1597); Physical Resistance (2578); Mechanical Jam (2983)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
Additional narrative: (b)(6).The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the mechanics swing fork and saw attachment were blocked.It was further reported that the mechanics were seized, jammed and heavy moving.Therefore, the reported condition was confirmed.The assignable root cause was determined to be wear from normal use and servicing.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that the saw attachment device was not oscillating.During an in-house engineering evaluation, it was observed that the mechanics swing fork and saw attachment were blocked.It was further reported that the mechanics were seized, jammed and heavy moving.It was not reported if the device was used in surgery or if there was patient involvement.It was not reported if there were any delays in a scheduled surgical procedure or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSCILL-SAW-ATTACHM F/532.001 532.010 532
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5092185
MDR Text Key26520980
Report Number3009450871-2015-13430
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Report Date 07/31/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2015
Date Manufacturer Received09/15/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/12/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-