Model Number 9-ASD-028
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Information (3190)
Event Date 09/03/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).The results of this investigation confirmed the aso met dimensional specifications when analyzed at sjm.A review of the device history record confirmed the occluder met all visual, dimensional, and functional specifications at the time it was manufactured, prior to shipment.There was no evidence to suggest there was an intrinsic defect in the occluder, and the cause for the embolization remains unknown.
 
Event Description
The patient's large atrial septal defect closure was attempted with a 26mm amplatzer septal occluder (aso) initially; however it was too small and was removed still attached to the delivery cable.A 28mm aso was implanted but embolized after it was released.The 28mm aso was percutaneously retrieved.A 30mm aso was attempted but was too small and removed still attached to the delivery cable.A 32mm aso was implanted successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key5099229
MDR Text Key26596573
Report Number2135147-2015-00109
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Report Date 09/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Expiration Date02/28/2019
Device Model Number9-ASD-028
Device Catalogue Number9-ASD-028
Device Lot Number4578818
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/23/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/24/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age15 YR
Patient Weight60
-
-