DENTSPLY TULSA DENTAL SPECIALTIES E3 MOTOR; CONTROLLER, FOOT, HANDPIECE AND CORD Back to Search Results
Catalog Number E3KIT
Device Problem Failure to Shut Off (2939)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Because this issue could possibly lead to file separation, which could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, it must be presumed that the issue would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device arrived in very bad conditions.The problem was due to the foot pedal; it was reworked by somebody, it is clear that on the switch there's a long scratch and a kind of "depression" in the plastic shell that create a shortcut inside.In fact removing the plastic shell the pedal returned to work normally.
 
Event Description
In this event it has been reported that an e3 motor is turning although device is switched off; no injury resulted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E3 MOTOR
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
DENTSPLY TULSA DENTAL SPECIALTIES
608 rolling hills drive
johnson city TN 37604
Manufacturer (Section G)
ADVANCED TECHNOLOGY RESEARCH (A.T.R.) S.R.L.
via san donato 1
pistoia, 51100
IT   51100
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key5105120
MDR Text Key26804319
Report Number9611053-2015-00041
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103653
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Report Date 08/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberE3KIT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/03/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/25/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/03/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-