The customer reported that they had questionable results for two patient samples tested for thyroid, bone, and fertility tests.When the samples were tested on the cobas e602 analyzer, these results did not match results when tested on an abbott architect i2000sr analyzer.The samples were also treated with heterophile blocking tubes (hbt) from scantibodies and lower results were found for several tests.Treatment with the hbt showed interference for all thyroid tests as well as other analytes.Turbidometric tests did not seem to be affected.The two patient samples were found to have erroneous results for free triiodothyronine (ft3), free thyroxine (ft4), thyrotropin (tsh), and thyroxine (t4).Only the results for ft3, ft4, and tsh were determined to be a reportable malfunction.The results were reported outside of the laboratory to the doctors for the first patient sample.For the second patient sample, the results were not reported outside of the laboratory and the requesting physician was informed of sample interference.This medwatch will cover ft4.Refer to the medwatch with (b)(6) for information referring to ft3.Refer to the medwatch with (b)(6) for information referring to tsh.The first patient sample initially resulted as 10.6 pmol/l for ft3, 56 pmol/l for ft4, and 0.621 mui/l for tsh.When the sample was treated with the hbt tube and tested on the cobas e602 analyzer, the sample resulted as 4.5 pmol/l for ft3, 20 pmol/l for ft4, and 1.02 mui/l for tsh.When tested on an abbott architect i2000sr analyzer, the sample resulted as 14 pmol/l for ft4 and 0.741 mui/l for tsh.When the sample was treated with the hbt tube and tested on the abbott architect i2000sr analyzer, the sample resulted as 14 pmol/l for t4 and 0.746 mui/l for tsh.The second patient sample, from a (b)(6), initially resulted as 13.4 pmol/l for ft3, 70 pmol/l for ft4, and 1.99 mui/l for tsh on (b)(6) 2015.When the sample was treated with the hbt tube and tested on the cobas e602 analyzer, the sample resulted as 4.6 pmol/l for ft3, 18 pmol/l for ft4, and 3.34 mui/l for tsh.When tested on an abbott architect i2000sr analyzer, the sample resulted as 14 pmol/l for t4 and 2.15 mui/l for tsh.When the sample was treated with the hbt tube and tested on the abbott architect i2000sr analyzer, the sample resulted as 15 pmol/l for t4 and 2.21 mui/l for tsh.The first patient received a thyroid ultrasonography based on the erroneous results, but both patients were not adversely affected.The cobas e602 analyzer serial number was (b)(4).
|