• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Catalog Number FEM07040
Device Problems Positioning Failure (1158); Premature Activation (1484)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/15/2015
Event Type  malfunction  
Manufacturer Narrative
The device history records are being reviewed.The event is currently under investigation.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant was unable to provide any patient details.
 
Event Description
It was reported that during implantation of the endovascular stent graft in a lower arm loop graft, the stent graft could not be deployed any further after being partially released.The device was retracted without issue and another stent graft was used to complete the procedure successfully.There was no reported patient injury.
 
Manufacturer Narrative
The lot history records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met all specifications prior to shipment.There is no indication for a manufacturing-related failure.The evaluation of the returned device confirmed the reported deployment failure.The stent graft was found to be partially released upon sample receipt and the outer sheath of the delivery system was found elongated.The condition of the device indicates the presence of increased friction during the attempt to deploy the stent graft.Potential contributing factors to the reported event have been considered.Previous investigations of similar complaints have been reviewed.This type of event may be associated with a difficult vessel anatomy leading to increased friction and subsequent partial deployment.In this case, no information regarding the anatomy was provided.No introducer sheath was used in this case which is considered a contributing factor to the reported event.Insufficient flushing of the device may be another contributing factor.In this case, it is unknown whether the device was flushed.On the basis of the information available, a definite root cause for the reported event could not be determined.The ifu states that the device must be flushed with sterile saline and recommends the use of an appropriate introducer sheath.Also the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure" and "do not kink the delivery catheter or use excessive force during delivery to the target lesion." also the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Type of Device
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG -9681442
wachhausstrasse 6
karlsruhe 76227
GM   76227
Manufacturer Contact
daniela mueller
wachhausstrasse 6
karlsruhe 76227
GM   76227
0497219445
MDR Report Key5110134
MDR Text Key27104965
Report Number9681442-2015-00167
Device Sequence Number1
Product Code PFV
UDI-Device Identifier04049519008431
UDI-Public(01)04049519008431(17)170911(10)ANYI1535
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130029/S003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Followup
Report Date 09/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/11/2017
Device Catalogue NumberFEM07040
Device Lot NumberANYI1535
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received11/18/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/29/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-