MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number 2AF284
Device Problem Insufficient Information (3190)
Patient Problems Cardiac Tamponade (2226); Pericardial Effusion (3271)
Event Date 09/16/2015
Event Type  Injury  
Manufacturer Narrative
This event occurred (b)(6) where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
 
Event Description
It was reported that during a cryoablation procedure, tamponade was suspected as the patient's blood pressure declined to the seventies.Pericardial effusion was found through observation with the echocardiogram, and drainage was performed.It was noted that the cause was unclear.The case was completed with cryo.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the data files were returned and analyzed.The data files did not show any system noticed on the date of the reported event.No product was returned for analysis.A clinical issue was encountered during the procedure; no product malfunction alleged.In conclusion, there was no indication of a product malfunction and no product returned.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER
Type of Device
PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland H9H 5 H3
CA  H9H 5H3
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5110470
MDR Text Key26986714
Report Number3002648230-2015-00323
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P100010/S015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Followup,Followup
Report Date 09/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/01/2016
Device Model Number2AF284
Device Catalogue Number2AF284
Device Lot Number12377
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/29/2015
Supplement Dates Manufacturer ReceivedNot provided
11/24/2015
Supplement Dates FDA Received11/30/2015
09/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
-
-