(b)(4) the results of the investigation are inconclusive since the device was not returned for analysis.Our investigation was limited to the review of the device history record, which showed that each manufacturing and inspection operation was performed and indicated complete in accordance with sjm specifications and procedures.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
A 28mm amplatzer septal occluder (aso) was selected for use.The patient's atrial septal defect (asd) was asymmetrical and measured 24mm x 27mm therefore the 28mm aso was implanted.The asd had a small rim which extended upward from the aorta.At implant, the edge of the aso came into minimal contact with the posterior wall of the left atrium but appeared to be stable and the position was considered acceptable although there was concern for potential erosion due to the position.Post-procedure, the patient was transferred to the ward and closely monitored.A transthoracic echo post-procedure noted that the aso had embolized to the mitral valve.The patient was not a candidate for percutaneous retrieval due to device size and emergency surgery to explant the aso and surgically repair the defect with a patch was undertaken.The patient was reported to be doing well post-operatively.
|