• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. (CS) SJM MECHANICAL HEART VALVE; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. (CS) SJM MECHANICAL HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 25A-101
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Code Available (3191)
Event Date 08/18/2015
Event Type  Injury  
Event Description
Prosthetic aortic stenosis was noted with this 25 mm sjm mechanical heart valve implanted at an unknown date.At the time of surgery, the valve appeared normal; however it is reported the valve was explanted and subvalvular stenosis was debrided.
 
Manufacturer Narrative
(b)(4).The results of this investigation concluded there was focal fibrin thrombus on the sewing cuff, and both leaflets were fully mobile with no resistance occurring during manipulation.There was no evidence of acute inflammation and a gram stain was negative for organisms.No evidence was found to suggest the cause of the thrombus formation was due to an intrinsic defect in the valve, as supported by the analysis performed.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MECHANICAL HEART VALVE
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC. (CS)
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. (CS)
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key5111915
MDR Text Key27052189
Report Number2648612-2015-00025
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Followup
Report Date 08/31/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number25A-101
Device Catalogue Number25A-101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/31/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age65 YR
-
-