ZOLL CIRCULATION AUTOPULSE® RESUSCITATION SYSTEM MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Positioning Failure (1158); No Display/Image (1183); Display or Visual Feedback Problem (1184); Visual Prompts will not Clear (2281); Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 08/26/2015
Event Type  malfunction  
Manufacturer Narrative
Zoll has not received the product in complaint.A supplemental report will be filed if and when the product is returned and investigation has been performed.
 
Event Description
It was reported that the leds on the user control panel of the autopulse platform were illuminated, when the device was powered on.However, the prompts on the lcd display were not visible.When the battery status check button was pressed on the battery that was used with this platform; the light emitting diodes illuminated, thereby providing an indication that the battery was ready for use.A training lifeband was installed.There was no report of any patient involvement.No additional details were provided.
 
Manufacturer Narrative
This supplemental is being submitted for mfr 3010617000-2015-00519, which is a duplicate of mfr 3010617000-2015-00553 that was submitted on 12/21/2015.The autopulse platform (s/n (b)(4)) was returned to zoll (b)(4) on 12/03/2015.Investigation results as follows: a visual inspection of the returned autopulse platform was performed and the short black cover was observed to be damaged.Visual inspection of the internal device showed no damages.A review of the archive was performed and no discrepancies were observed.The platform was functional tested and the autopulse the reported complaint of the device does not turn on was confirmed.Further investigation indicated that the processor board and the cable from the load cell amplifier to the processor board were defective.Replacing the parts remedied the ua7.The platform passed the load cell characterization test.Based on the visual inspection, the short black cover was replaced.Additionally, the power distribution board was also replaced to remedy the reported complaint.In summary, the customer's reported complaint of the platform not turning on and not leds illuminating was confirmed during functional testing.After replacement of the part identified during investigation, the platform passed all final functional testing criteria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION SYSTEM MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
joy patel
2000 ringwood ave,
san jose, CA 95131
4084192957
MDR Report Key5114226
MDR Text Key27145166
Report Number3010617000-2015-00519
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Followup
Report Date 09/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received12/04/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-