• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 06437281190
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/16/2015
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer questioned results from 2 patient samples tested for thyrotropin (tsh), free thyroxine (ft4) and free triiodothyronine (ft3).The customer provided the samples for investigation.Of the data provided, erroneous ft4 results were identified between the customer's e602 analyzer, a centaur analyzer, an e 170 analyzer used at the investigation site and an e411 analyzer used at the investigation site.The date of tests performed at the customer site is not known.It is not known if erroneous results were reported outside of the laboratory.Refer to the attachment to the medwatch for patient results.No adverse event was reported.The e 170 analyzer serial number was (b)(4).The e411 analyzer serial number was (b)(4).The ft4 reagent lot number used at the investigation site was 180539 with an expiration date of 10/31/2015.The serial number for the customer's e602 analyzer is not known.A specific root cause could not be identified.Based on the available data, a general reagent issue could be excluded.When comparing values from different types of analyzers, variances can be expected.For thyroid parameters, age, gender and other patient characteristics should be considered when comparing values.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5117590
MDR Text Key27248890
Report Number1823260-2015-04268
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Report Date 10/01/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/01/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number06437281190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Event Location Laboratory
Date Manufacturer Received09/18/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-