Device Problem
Battery Problem (2885)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/09/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Ge healthcare¿s investigation into the reported occurrence is still ongoing.A follow-up report will be issued when the investigation has been completed.No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.
|
|
Event Description
|
The hospital reported that, during a preoperative checkout, the unit did not switch to battery backup during ac power loss.There was no report of patient involvement.
|
|
Manufacturer Narrative
|
A ge healthcare service representative performed a checkout of the equipment and confirmed the reported complaint.The battery was replaced, and the unit was returned to service.
|
|