It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the roller pump does not display the tubing size or the flow.The device was not changed out, as they continued to use for the case.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.Per the clinical review on 15-sep-2015: according to the lead perfusionist (ccp), the issue was noticed during cardiopulmonary bypass that the roller pump would intermittently not display the tubing size and flow.They continued to use the pump for the remainder of the case, but the flow was unknown.There was no impact to the patient as a result of the missing data.The case was completed successfully, without delay and without associated blood loss.There was no harm observed.
|