The case states that the facility's dsd-201 automated endoscope reprocessor (aer) was set to the wrong program.The aer was set to have only 2 rinses when rapicide opa/28 instructions require 3 and thus was not adequately rinsing endoscopes during reprocessing cycles.The facility called medivators technical service to see if their aer was set up properly and it was discovered that the machine was set to the wrong program.Medivators technical service assisted the facility in changing the aer to the correct program.It is unknown how long the facility was reprocessing scopes with the wrong number of rinses.Improper rinsing could leave residual disinfectant on the scope, therefore potential handler and patient chemical exposure symptoms such as irritation or chemical colitis.To date, there have been no reports of illness or injury.This complaint will continue to be monitored within the medivators complaint handling system.
|