• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+; CT STATUS+

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+; CT STATUS+ Back to Search Results
Catalog Number 10379675
Device Problems Smoking (1585); Device Displays Incorrect Message (2591); Sparking (2595)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/08/2015
Event Type  malfunction  
Manufacturer Narrative
Siemens is in process of investigating customer returned materials.The cause for the event is unknown at this moment.
 
Event Description
Customer reported that connector base started smoking and sparking and it caused the instrument to give error (e02) code.There was no report of injury due to this event.
 
Manufacturer Narrative
Siemens hsc team investigated customer returned instrument and root cause was found to be defective choke components on the main board.This is a known issue, and corrective action has been implemented from serial # (b)(4).Resolution was alternative supplier of component.No further investigation is required.Customer was sent a replacement instrument.Upon follow up, customer confirmed that the new instrument has been running well, and she needs nothing else.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+
Type of Device
CT STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key5131569
MDR Text Key27967322
Report Number1217157-2015-00143
Device Sequence Number1
Product Code JIL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Type of Report Followup
Report Date 11/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number10379675
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/07/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/16/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-