Model Number 5392
Device Problems Mechanical Problem (1384); Nonstandard Device (1420); Device Displays Incorrect Message (2591); Power Problem (3010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.Product event summary: analysis confirmed the reported event, the keypad was out of electrical specification.It was also noted that the output connector assembly wire insulation was pinched but the insulation was not compromised, the display wire insulation was pinched but the insulation was not compromised, and it was noted in the log that one stuck button was detected (on/off) and one stuck button was recovered.The device shutdown during testing which caused a system crash, giving no test results.After the main printed circuit board was realigned, test results were available which failed due to an error and on/off button failure.(b)(4).
 
Event Description
It was reported that the external pulse generator did not turn on without the on/off button being pressed repeatedly and that once it did turn on it generated an error message "button press detected." the generator was returned for service.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5133588
MDR Text Key28067383
Report Number3004593495-2015-00215
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Remedial Action Recall
Report Date 07/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received07/14/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1356-2015