Model Number GF-UE160-AL5
Device Problem Insufficient Information (3190)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/09/2015
Event Type  malfunction  
Manufacturer Narrative
The device referenced in this report has not been returned to olympus for evaluation.Therefore, the exact cause could not be determined at this time.The manufacturing history was reviewed, with no irregularities related to this problem noted.If additional information becomes available at a later time, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that bacillus subtilis was detected from the biopsy channel's rinsing water, and enterococcus faecalis and staphylococcus were detected from the biopsy channel's suction liquid while the culture examination was conducted by the user facility.There was no patient harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC GASTROVIDEOSCOPE
Type of Device
ULTRASONIC GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951
ishikawa-cho
hachioji-shi, tokyo 192-8 507
JA  192-8507
Manufacturer Contact
susumu nishina
2951
ishikawa-cho
hachioji-shi, tokyo 192-8-507
JA   192-8507
6425177
MDR Report Key5133936
MDR Text Key28072222
Report Number8010047-2015-00954
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K#:K051541
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Service Personnel
Report Date 09/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGF-UE160-AL5
Device Catalogue NumberN/A
Device Lot NumberN/A
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/15/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/07/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-