MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH SIZE #6 4-IN-1 CUTTING BLOCK CAPTURED ASSY TRIA. EXP. INSTR.; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH SIZE #6 4-IN-1 CUTTING BLOCK CAPTURED ASSY TRIA. EXP. INSTR.; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 6541-1-706E
Device Problems Break (1069); Detachment Of Device Component (1104); Manufacturing, Packaging or Shipping Problem (2975)
Patient Problems Bone Fracture(s) (1870); Injury (2348); No Known Impact Or Consequence To Patient (2692)
Event Date 09/18/2015
Event Type  malfunction  
Manufacturer Narrative
When completed, the investigation results will be submitted in a supplemental report.
 
Event Description
It was reported that during a right primary knee surgery, upon removal of the #6 4 in 1 cutting block, the peg broke off and was left in the patient's bone.The piece was immediately retrieved without any delay and the case completed without any other events.
 
Manufacturer Narrative
An event regarding pin dissociation of a triathlon 4:1 express cutting guide was reported.The event was confirmed.Method and results: device evaluation and results: inspection of the returned device confirmed the pin had dissociated from the device body.Medical records received and evaluation: not performed as there was no indication that patient factors contributed to the reported event.Device history review: all devices accepted into final stock conformed to specification.This review confirmed the device was manufactured prior to capa implementation.Complaint history review: there have been no similar previous reported events for this lot id.Conclusions: the investigation concluded that the fixation peg disassociating from the triathlon 4:1 express cutting block was caused by a manufacturing nonconformance.It was concluded that the supplier, (b)(4), had not performed the required press fit operation between the peg and block which led to the pin coming out of the assembly.Stryker reserves the right to re-evaluate this investigation if additional relevant information becomes available.
 
Event Description
It was reported that during a right primary knee surgery, upon removal of the #6 4 in 1 cutting block, the peg broke off and was left in the patient's bone.The piece was immediately retrieved without any delay and the case completed without any other events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIZE #6 4-IN-1 CUTTING BLOCK CAPTURED ASSY TRIA. EXP. INSTR.
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla navedo
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5138371
MDR Text Key28077080
Report Number0002249697-2015-03275
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123486
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Followup
Report Date 09/18/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6541-1-706E
Device Lot NumberSB3M86
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/08/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/21/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age53 YR
Patient Weight105
-
-