STRYKER ORTHOPAEDICS-MAHWAH SIZE #6 4-IN-1 CUTTING BLOCK CAPTURED ASSY TRIA. EXP. INSTR.; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
Catalog Number 6541-1-706E |
Device Problems
Break (1069); Detachment Of Device Component (1104); Manufacturing, Packaging or Shipping Problem (2975)
|
Patient Problems
Bone Fracture(s) (1870); Injury (2348); No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/18/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
When completed, the investigation results will be submitted in a supplemental report.
|
|
Event Description
|
It was reported that during a right primary knee surgery, upon removal of the #6 4 in 1 cutting block, the peg broke off and was left in the patient's bone.The piece was immediately retrieved without any delay and the case completed without any other events.
|
|
Manufacturer Narrative
|
An event regarding pin dissociation of a triathlon 4:1 express cutting guide was reported.The event was confirmed.Method and results: device evaluation and results: inspection of the returned device confirmed the pin had dissociated from the device body.Medical records received and evaluation: not performed as there was no indication that patient factors contributed to the reported event.Device history review: all devices accepted into final stock conformed to specification.This review confirmed the device was manufactured prior to capa implementation.Complaint history review: there have been no similar previous reported events for this lot id.Conclusions: the investigation concluded that the fixation peg disassociating from the triathlon 4:1 express cutting block was caused by a manufacturing nonconformance.It was concluded that the supplier, (b)(4), had not performed the required press fit operation between the peg and block which led to the pin coming out of the assembly.Stryker reserves the right to re-evaluate this investigation if additional relevant information becomes available.
|
|
Event Description
|
It was reported that during a right primary knee surgery, upon removal of the #6 4 in 1 cutting block, the peg broke off and was left in the patient's bone.The piece was immediately retrieved without any delay and the case completed without any other events.
|
|
Search Alerts/Recalls
|
|
|