Analysis of the instrument and instrument data indicate that the cause for the falsely elevated hb1c results is user error.The account failed to input the correct lot specific scaler values.The siemens healthcare diagnostics customer care center performed reagent calibration data analysis and informed the customer of their incorrect scaler entry.The issue was resolved by entering correct scaler values, recalibrating, and verifying with qc.The hb1c flex reagent cartridge df105a instructions for use clearly state: hb1c requires lot-specific scalers which must be entered in the calibration set up screen, prior to calibration.The scaler values are provided on the flex reagent cartridge carton.These scalers are applied to all qc and patient results to maintain accuracy.Failure to enter the lot-specific scalers will cause inaccurate results.The instrument is performing within specifications.No further evaluation of the device is required.
|