MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Pericardial Effusion (3271)
Event Date 09/17/2015
Event Type  Injury  
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.(b)(4).
 
Event Description
It was reported that during a cryoablation procedure, a system notice was received indicating that the safety system detected a compromised outer vacuum.The catheter and coaxial umbilical cable were changed which resolved this issue.Additionally, a system notice was received indicating that the system detected a temperature drop during the system flush, and the system flush was stopped.The cryoconsole unit was rebooted without resolve.It was confirmed that the healthcare professionals were using a saline/contrast flush and that the temperature was reading 31.The healthcare professional was instructed to hold off on the flush and they were able to pass the integrity test.The cryoconsole unit appeared to be functioning normally at that time of the first ablation.The case was then aborted due to a pericardial effusion noted on intracardiac ultrasound.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: data file analysis was performed.The sheath was not returned for analysis.An adverse event was encountered during the procedure.In conclusion, an adverse event occurred during the procedure (pericardial effusion) and the sheath was not returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland H9H 5 H3
CA  H9H 5H3
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5144906
MDR Text Key28042786
Report Number3002648230-2015-00355
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Followup
Report Date 09/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/17/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
-
-