HOLOGIC, INC. AQUILEX; INSUFFLATOR, HYSTEROSCOPIC Back to Search Results
Device Problems Mechanical Problem (1384); Improper Flow or Infusion (2954)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/04/2015
Event Type  malfunction  
Event Description
Resectoscope was inserted with a wire loop.Fibroid was completely resected with the wire loop using 100 watts coagulation and cutting current.The fibroid bed and part of the uterus was successfully ablated; however, the fluid machine which they were using registered over 1000 ml, although by their count they only had 700 ml deficit.The machine shut off, it would not allow any more fluid.The procedure was then terminated having partially ablated via endometrial lining.The resectoscope was removed.The patient was taken to the recovery room in stable condition.Manager of surgery will meet with manufacturer's representative to discuss issues we were having with this machine during the case.This was a new machine and the fluid flow was noted to be fast.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUILEX
Type of Device
INSUFFLATOR, HYSTEROSCOPIC
Manufacturer (Section D)
HOLOGIC, INC.
250 campus drive
marlborough, MA 01752
MDR Report Key5147793
MDR Text Key28255775
Report Number5147793
Device Sequence Number1
Product Code HIG
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other
Report Date 09/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date10/31/2016
Was the Report Sent to FDA? Yes
Date Report Sent to FDA09/17/2015
Event Location Hospital
Date Report to Manufacturer09/17/2015
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/14/2015
Type of Device Usage N
Patient Sequence Number1
Patient Age54 YR
-
-