Catalog Number 06437281190
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/26/2015
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer received questionable triiodothyronine (t3), free thyroxine (ft4) and thyrotropin (tsh) results for one patient from cobas e602 analyzer serial number (b)(4) when compared to a vitros eci analyzer.Refer to the attachment to the medwatch for all patient data.(b)(4).The results were reported to the doctor.Based on the results from the cobas e602, treatment with strumazol was initiated to treat elevated hormonal production.The strumazol treatment was terminated on (b)(6) 2015.No adverse effects were reported for the patient.
 
Manufacturer Narrative
Samples from the patient were submitted for investigation and no interfering factor to streptavidin could be identified.
 
Manufacturer Narrative
Further testing of the samples revealed an interfering factor was present in the sample which most likely caused the issue.Product labeling documents this interference.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5150964
MDR Text Key28305298
Report Number1823260-2015-04336
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Followup,Followup
Report Date 12/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number06437281190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Event Location Laboratory
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/14/2015
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received10/28/2015
12/02/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
LANETTECREAM I FNA
-
-