Nexwave, (b)(4), evaluated on (b)(4) 2015.No problem found on the actual device.Ran unit in ifc mode for 10 minutes w/out any abnormal heating issue.Ran the unit for an additional 15 minutes, which the ifc was hooked up to the lbg, again, no issue.Lbg; supply item, was evaluated as well, because the patient claimed that the electrodes wouldn't stay fastened, regardless of the lbg used.The evaluation revealed there was no problem found.Engineering technician suggests that there was poor attachment of electrodes causing poor adherence.Patient error.
|