ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3788
Device Problems Nonstandard Device (1420); Self-Activation or Keying (1557)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Manufacturer Narrative
(b)(4).This ipg serial number was included in a field advisory.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.(b)(4).
 
Event Description
It was reported the patient's scs ipg was turning on by itself without any magnetic interference ( date of event unknown).As a result, surgical intervention was taken where the scs system was explanted.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key5153445
MDR Text Key28408115
Report Number1627487-2015-20570
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Remedial Action Recall
Type of Report Followup
Report Date 11/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2013
Device Model Number3788
Device Lot Number3333249
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/15/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/23/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/03/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-002-R
Patient Sequence Number1
Patient Outcome(s) Other;
-
-