The companion 2 driver was not supporting a patient.The customer reported that the companion 2 driver was exhibiting irregular pressure waveforms.This alleged failure mode poses a low risk to a patient because the issue was observed when the companion 2 driver was not supporting a patient.The companion 2 driver was returned to syncardia for evaluation.The patient file was copied and reviewed, revealing no anomalies.Visual inspection of the driver's external components revealed no anomalies.During investigation testing, the reported irregular pressure waveform was confirmed.The root cause of the irregular pressure waveform was a malfunction of the right electronic pressure regulator.The right electronic pressure regulator was taken out of service and replaced.The issue with the electronic pressure regulators in relation to irregular pressure waveforms regulator is being investigated in a capa (corrective or preventive action).The capa will document the investigation including, but not limited to, potential root cause and corrective actions associated with compressor cycling because of a malfunction of the manual pressure regulator.A small timing difference between the left and right pressure waveforms was also observed during failure investigation testing, which was caused by a malfunction of the right pilot valve.The right pilot valve was taken out of service and replaced.These failure modes posed a low risk to a patient because the driver was not supporting a patient at the time of the reported issue, and they would not have prevented the driver from performing its life-sustaining functions.The companion 2 driver was serviced and passed all functional and performance testing before being placed into finished goods.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
|