AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; CARDIAC OCCLUSION DEVICE Back to Search Results
Model Number 9-ASD-032
Device Problems Migration or Expulsion of Device (1395); Device Dislodged or Dislocated (2923)
Patient Problem Embolism (1829)
Event Date 07/16/2015
Event Type  Injury  
Manufacturer Narrative
Gtin: (b)(4).
 
Event Description
One day post-implant, a 32mm amplatzer septal occluder (aso) was found embolized in this asymptomatic patient.Surgery was required to retrieve the aso and close the atrial septal defect.The patient was reported to be stable.
 
Manufacturer Narrative
The results of this investigation confirmed the amplatzer septal occluder met dimensional specifications when analyzed at sjm.A review of the device history record confirmed the occluder met all visual, dimensional, and functional specifications at the time it was manufactured, prior to shipment.There was no evidence to suggest there was an intrinsic defect in the occluder, and the cause for the embolization remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key5161472
MDR Text Key28720369
Report Number2135147-2015-00124
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Followup
Report Date 09/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Expiration Date11/30/2018
Device Model Number9-ASD-032
Device Catalogue Number9-ASD-032
Device Lot Number4438528
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/20/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/29/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/30/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-