ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 06437281190
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/02/2015
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer received questionable thyroid results for two patient samples from a cobas e602 analyzer.The specific date of testing was not provided.The samples were submitted for investigation and were tested on analytical e module, cobas e411, and centaur analyzers.Of the data provided, only the results for free triiodothyronine (ft3) and free thyroxine (ft4) were discrepant.Refer to the attachment to the medwatch for all patient data.Refer to the medwatch with (b)(6) for the ft3 assay.Information concerning if any result was reported outside the laboratory was requested, but was not provided.No adverse event was reported.A specific root cause could not be identified.Additional information for further investigation was requested but was not provided.From the information provided, a general reagent issue was not suspected.Differences in values between the roche and siemens analyzers could be due to the different setups of all assays, the antibodies used and the variances in reference methods.The results of all thyroid parameters vary in function of age, gender and other population characteristics which should be taken into account when analyzing the results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5163277
MDR Text Key28806931
Report Number1823260-2015-04368
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Report Date 10/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number06437281190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Event Location Laboratory
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/20/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-