• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 980 VENTILATOR; CONTINUOUS VENTILATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 980 VENTILATOR; CONTINUOUS VENTILATOR Back to Search Results
Model Number 980
Device Problems Device Emits Odor (1425); Smoking (1585)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/07/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).The service engineer (se) evaluated the ventilator, and verified the customer reported malfunction.The se replaced the breath delivery power controller, printed circuit board assembly (pcba), to resolve the issue.
 
Event Description
It was reported that after a month of patient use, a ventilator was removed from a patient and placed in standby mode.The patient and device were then moved to a new location.During set-up in the new location, smoke was observed coming from the area around the exhalation filter and a burning smell was detected.The patient was transferred to another ventilator without harm.
 
Manufacturer Narrative
The device was repaired and the reported issue was isolated to interface between the device and the failed component.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Type of Device
CONTINUOUS VENTILATOR
Manufacturer (Section D)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI 
Manufacturer (Section G)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI  
Manufacturer Contact
thom mcnamara
15 hampshire street
mansfield, MA 02048
5084524811
MDR Report Key5169946
MDR Text Key29006581
Report Number8020893-2015-01269
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131252
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/22/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number980
Device Catalogue Number980
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received10/08/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/27/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-