Model Number 980 |
Device Problems
Device Emits Odor (1425); Smoking (1585)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/07/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).The service engineer (se) evaluated the ventilator, and verified the customer reported malfunction.The se replaced the breath delivery power controller, printed circuit board assembly (pcba), to resolve the issue.
|
|
Event Description
|
It was reported that after a month of patient use, a ventilator was removed from a patient and placed in standby mode.The patient and device were then moved to a new location.During set-up in the new location, smoke was observed coming from the area around the exhalation filter and a burning smell was detected.The patient was transferred to another ventilator without harm.
|
|
Manufacturer Narrative
|
The device was repaired and the reported issue was isolated to interface between the device and the failed component.If information is provided in the future, a supplemental report will be issued.
|
|