This freedom ac power supply was not in patient use.The freedom a/c power supply is a component that enables the freedom driver to be plugged into an external power source.The customer reported that green connector on the freedom a/c power supply would not connect to the green receptacle on the freedom power adaptor.The freedom ac power supply was returned to syncardia for evaluation.Visual inspection of the freedom hospital ac power supply revealed that the connector cover was not in the correct orientation.This is the root cause of the customer-reported inability to connect the ac power supply to a freedom power adaptor.It is likely that the cable connector cover was installed incorrectly at syncardia during incoming inspection.During incoming inspection at syncardia, inspectors perform a verification to ensure the connector's interior threads are properly mated with adhesive.The verification requires inspectors to remove the connector cover.Incoming inspection requirements were revised to clearly identify and verify the correct alignment of the connector cover.This failure mode poses a low risk to a patient because the freedom hospital ac power supply was not in use by a patient at the time of the customer-reported issue.In addition, it would not prevent a freedom driver from performing its life-sustaining functions.The freedom driver is equipped with redundant power sources, including an additional ac power supply and multiple rechargeable freedom onboard batteries.The freedom hospital ac power supply cable was disassembled, the connector key slot was aligned with the connector cover key tab and the correct mating configuration was obtained.The freedom hospital ac power supply was reassembled, serviced, and passed all functional and performance testing before being placed into finished goods.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
|