• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE 10MM, 33CM PEEK MONOPOLAR HANDLE; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE 10MM, 33CM PEEK MONOPOLAR HANDLE; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED Back to Search Results
Catalog Number 0250080618
Device Problems Crack (1135); Peeled/Delaminated (1454)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/29/2015
Event Type  malfunction  
Manufacturer Narrative
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
 
Event Description
It was reported the insulation at the tip of the shaft is cracked/peeling.
 
Manufacturer Narrative
"the following repair and service diagnostic codes were identified: cracked/peeling insulation at tip of shaft.Replaced handle.The following was observed: cracked / peeling insulation at tip of shaft.The handle was replaced.This does confirm the alleged failure mode of: insulation damage.Probable root cause: "poor autoclave reliability.Incorrect sterilization/reprocessing procedure.Handling procedures.Contact forces.Product used beyond defined useful life." the product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
 
Event Description
It was reported the insulation at the tip of the shaft is cracked/peeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10MM, 33CM PEEK MONOPOLAR HANDLE
Type of Device
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
kimberly lynch
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key5178879
MDR Text Key29931236
Report Number0002936485-2015-00954
Device Sequence Number1
Product Code OCZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250080618
Device Lot Number¿1343724¿
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/29/2015
Initial Date FDA Received10/26/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/15/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-