Product event summary: the flexcath advance sheath 4fc12 / (b)(4) was visually inspected and functionally tested.Upon visual inspection of sheath, results did not show any detachment of the stopcock distal end luer from the side port tube proximal end.Both the sheath's distal tip and dilator tip were intact.There were no fractures, breaks, or kinks noticed.No teflon delamination was noticed at the tip of the sheath shaft.The sheath tip was atraumatic and with no sharp edges.Upon functional testing of the sheath, results showed that the deflection worked as per specification.In conclusion, the reported ¿stopcock detachment¿ issue has not been confirmed through visual inspection and testing.The sheath passed the returned product inspection as per specification.
|