Model Number 4FC12
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/30/2015
Event Type  malfunction  
Manufacturer Narrative
 
Event Description
It was reported that during a cryoablation procedure, the stopcock detached from the sheath.The sheath was replaced with a new one and the procedure was completed with cryo.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the flexcath advance sheath 4fc12 / (b)(4) was visually inspected and functionally tested.Upon visual inspection of sheath, results did not show any detachment of the stopcock distal end luer from the side port tube proximal end.Both the sheath's distal tip and dilator tip were intact.There were no fractures, breaks, or kinks noticed.No teflon delamination was noticed at the tip of the sheath shaft.The sheath tip was atraumatic and with no sharp edges.Upon functional testing of the sheath, results showed that the deflection worked as per specification.In conclusion, the reported ¿stopcock detachment¿ issue has not been confirmed through visual inspection and testing.The sheath passed the returned product inspection as per specification.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland H9H 5 H3
CA  H9H 5H3
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5180818
MDR Text Key30038163
Report Number3002648230-2015-00374
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 12/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/07/2017
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number87759
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/09/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/08/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-