MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. UNKNOWN DEPUY ELITE FEMORAL STEM; HIP FEMORAL STEM/SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. UNKNOWN DEPUY ELITE FEMORAL STEM; HIP FEMORAL STEM/SLEEVE Back to Search Results
Catalog Number UNK-HIP
Device Problem Loss of or Failure to Bond (1068)
Patient Problem Pain (1994)
Event Date 10/22/2015
Event Type  Injury  
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.See section for any product information received.Depuy synthes has been informed that the catalog number and lot number is not available.
 
Event Description
Patient was revised to address pain and femoral stem loosening at the cement to implant interface.Cement used at the time of implant is unknown.
 
Manufacturer Narrative
No device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review was not possible because the required lot code(s) was not provided.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN DEPUY ELITE FEMORAL STEM
Type of Device
HIP FEMORAL STEM/SLEEVE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key5182250
MDR Text Key29403590
Report Number1818910-2015-33878
Device Sequence Number1
Product Code KXA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK-HIP
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/22/2015
Initial Date FDA Received10/27/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/05/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age83 YR
Patient Weight84
-
-