Dhr review: a full review of the device history record for this device has been carried out with no anomalies identified; product was manufactured and released for sale in accordance with specifications.There is no information to suggest that the device has not met the final release criteria.Video review: additionally review of an in process video taken after loading of the main body and proximal extender devices was reviewed.This video acts as a record that the device meets the inspection requirements for a loaded implant.This video confirmed that the devices were packed into the delivery systems in accordance to company work instruction 211, inspection of packed sg-hbb and sg-hpe devices.Case review: original evar was performed on (b)(6) 2015.The main body was implanted without issue.However, during the procedure, the physician stated that it was not possible to cannulate the contralateral side.The left common iliac artery was very tortuous and it was not possible to straighten with a stiff guide wire.An 8f flexor sheath and a guide wire were used to cross the angulated section.It is believed that the dissection may have been caused whilst performing this part of the procedure.From a review of the patient's sizing sheet, it is clear that the left common iliac was aneurysmal (up to 30mm) with partial calcification and tortuous vessels.It is possible that the dissection was caused when attempting to maneuver through the common iliac.A convertor device was implanted following identification of the dissection in an attempt to stop the loss of blood from the dissection.The converter device has not caused or contributed to this event and was implanted in an attempt to resolve the issue and stabilize the patient.It has been reported that the patient had poor cardiac constitution.There is no information to suggest malfunction of the implant.The dhr review and video reviews are acceptable, and therefore confirm that all inspection and manufacturing activities met specifications.The stent graft has performed as intended.No further investigation is required.Ifu review: potential adverse events related to the procedure or device malfunction include, but are not limited to: death, cardiac complications e.G.Myocardial infarction, blood loss warnings and precautions state: care must be taken in those patients where the aneurysm proximal neck, distal landing zones, or iliac arteries are calcified and tortuous.Risk analysis: this event falls outside of lombard medical's risk analysis which states that adverse events associated with all surgery will not be covered by the analysis.Eg: bowel, stroke, respiratory.It has been reported that the patient was in poor cardiac constitution and this event is related to the surgery performed.The current event rate for patient death reported within lombard medical complaint system is 0.16%, which is within clinical expectations.Conclusion: the risk / benefit remains acceptable however lombard medical will continue to track and trend in accordance with quality system procedures.Lombard medical now intends to close this complaint.
|