Submit date: 01/25/2016.The device history record was reviewed and indicated that the product was released accomplishing all quality standards.There were no samples received for evaluation.Because a sample was not available for evaluation, a root cause analysis could not be conducted to determine the root cause of this reported issue.If samples are received at a later date, this complaint will be re-opened and the investigation continued.A possible root cause can be due to over clamping the device.A corrective action is not required at this time.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection, leak testing and visual acceptance sampling) are in place to prevent non-conforming product from leaving the manufacturing operations.As per procedure, manufacturing performs 100% leak testing and visual inspection at the final stage of production, which would identify this issue in the catheter assembly.This complaint will be used for tracking and trending purposes.
|