Catalog Number 5100009000
Device Problem Device Markings/Labelling Problem (2911)
Patient Problem No Patient Involvement (2645)
Event Date 10/14/2015
Event Type  malfunction  
Manufacturer Narrative
The handswitch is not a repairable device and will not be returned to the user facility.
 
Event Description
The handswitch was returned to stryker instruments for service.During functional testing by a service technician, it was found that the run/safe etchings are not legible, presenting a potential for the device to inadvertently be activated.No patient involvement, no delay, no medical intervention, and no adverse consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL HANDSWITCH
Type of Device
UNIT, OPERATIVE DENTAL
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key5203955
MDR Text Key30464364
Report Number0001811755-2015-04004
Device Sequence Number1
Product Code EIA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032117
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5100009000
Device Lot Number11019
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2015
Initial Date Manufacturer Received 10/14/2015
Initial Date FDA Received11/05/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/19/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-