DEPUY SYNTHES POWER TOOLS BATTERY OSCILLATOR; MOTOR, SURGICAL INSTRUMENT, AC- POWERED Back to Search Results
Catalog Number 530.610
Device Problem Device Stops Intermittently (1599)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/26/2015
Event Type  malfunction  
Manufacturer Narrative
As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a supplemental medwatch will be filed as appropriate.
 
Event Description
This is report 1 of 2 for the same event.It was reported that during an unspecified surgery, it was observed that the battery oscillator device was operating intermittently.An identical spare device was used; however, yielded the same results.A third identical spare device was used to complete the surgery.There were no delays reported to the surgical procedure.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device has intermittent operation and vibration.It was determined that the electronic control unit was damaged, the electric motor had extreme vibration, the o-rings were damaged, and the bearings and saw-head were worn out.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY OSCILLATOR
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC- POWERED
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5204770
MDR Text Key30806021
Report Number3009450871-2015-14112
Device Sequence Number1
Product Code GEY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/26/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.610
Device Lot Number4283
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/26/2015
Initial Date FDA Received11/05/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/16/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/21/2010
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-