AV-TEMECULA-CT SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING STENT SYSTEM Back to Search Results
Catalog Number S-55-060-120-P6
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 10/16/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that during unpacking of a 5.5x60 mm supera stent system a hair was noted to be embedded in the tray underneath the device protective tubing of the device.The device packaging was noted to be completely sealed prior to opening the device.The device was not used and there was no patient involvement.A new same sized supera stent was used to complete the procedure.There was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4)./ evaluation summary: unique device identifier (udi): (b)(4).The device was returned for evaluation.The reported hair inside the package was able to be confirmed.A review of the electronic lot history record (elhr) for the reported lot revealed no nonconforming material records (ncmrs), indicating all lot release testing met specifications.A query of the complaint-handling database indicated there were no similar incidents reported from this lot for this issue.In this case, there were no untoward patient effects since the device was not used on a patient and the issue was detectable prior to the procedure.Based on the information reviewed for this lot, it was determined that this was an isolated incident and not representative of the entire lot.This type of reported event will continue to be monitored per local quality system procedures.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Type of Device
SELF EXPANDING STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
WEBSTER, TX USA REG#3005325609
abbott vascular
26531 ynez rd.
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key5207931
MDR Text Key30843604
Report Number2024168-2015-06632
Device Sequence Number1
Product Code NIP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 02/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2016
Device Catalogue NumberS-55-060-120-P6
Device Lot Number02399058
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer11/02/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/16/2015
Initial Date FDA Received11/06/2015
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received01/19/2016
02/04/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial